| Operation: |
With touch panel LCD; Controllable via network (with IE6 web browser) |
|---|---|
| Incubator volume: | 460L |
| Maximum workpiece weight: | 15kg (up to 5kg accuracy guaranteed) |
| Temperature control: | Direct control via heater panel Range: room temperature +5°C to 40°C (max.), 0.1°C increments |
| Humidity control: | Via aerosol spray humidifier Range: 70% to 95%, 1% increments |
| CO2 intensity control: | CO2 supply: by external CO2 gas cylinder connection Range: 0% to 20%, 0.1% increments |
| Number of vessels mountable on each tier (max.): | Flask: 25cm2 x 2, 75cm2 x 1 Dish: Ø35mm x 5, Ø60mm x 2, Ø100mm x 1 Well plate: 4-well x 1, 6-well x 1, 12-well x 1, 24-well x 1 and 96-well x 1 |
| Specimen supply: | With dedicated carrier via access gate |
| Specimen storage rack: | 3 rows x 10 tiers |
| Macro observation: | Image capture of whole vessel with dedicated camera (bird's-eye view) Camera head: color CCD camera (1280 x 960 pixels) Brightfield: backlight illumination |
| Micro observation: | Magnification: 2x, 4x, 10x, 20x, 40x Objective: Plan Fluor DL/Plan Fluor ADL series Camera head: 2/3-inch cooled CCD camera (1M pixels, 15fps) Phase contrast: high-intensity red LED illumination, automatic phase ring changeover Epi-fluorescence (option): LED V/B/G illumination, 3 filter blocks mountable |
| Stage travel: | XY: 120 x 120mm (max. resolution: 8µm) Z: 6mm (max. resolution: 0.03125µm) |
| Sample focusing: |
AF focus point automatic detection 15 consecutive Z-axis scans manually possible both above and below focus point |
| Observation: | Via PC monitor |
| Power source: |
Power capacity: 100, 115, 230VAC ±10% 1300VA 50/60Hz Frequency: 50/60Hz |
| Weight: | Approx. 470kg |
| Operating environment: | Temperature: 15 - 28°C Humidity: Max. 60% relative humidity (noncondensing) |