Nikon Instruments / Prodotti / Sistemi di Microscopia / Microscopi Confocali / A1R-A1 Confocale

Nikon A1R-A1 Confocal Microscope System

A1R-A1 Specifications

A1R

Input/output port: Laser input port: 3 (FC x2, direct x1)
 Signal output port: 4 (SMA x2, FC x1, VAAS x1)
Laser
Wavelength and power: 405LD: max. 38mW, Multi-Ar (457/488/514): max. 65mW, 488DPSS: max. 75mW, 561DPSS: max. 25mW, 543HeNe: max.
 1mW, 638LD: max. 10mW
 440LD: max. 15mW (available as option)
Modulation: Method: AO (Acousto) device or drive current control
 Control: power control for each wavelength, Return mask, ROI exposure control
Laser unit: Standard: LU4 4-laser unit
 Optional: C-LU3EX 3-laser unit EX
Standard fluorescence detector
Detector: 4 PMT
Filter cube: 6 filter cubes commonly used for a microscope mountable on each of three filter wheels
 Recommended wavelengths: 450/50, 482/35, 525/25, 595/50, 700/75, 540/30, 515/30, 585/65
Diascopic detector
Detector: PMT
Scanning head
Scanning: Scanning range: square inscribed in a ø18mm circle
 Standard image acquisition
 Scanner: non-resonant scanner x2
 Pixel size: max. 4096 x 4096 pixels
 Scanning speed: 4 fps (512 x 512 pixels)
 Zoom: 1-1000x continuously variable
 Scanning mode: X-Y, XY rotation, Free line, Line Z
 High-speed image acquisition
 Scanner: resonant scanner (X-axis, resonance frequency 7.8kHz),
 non-resonant scanner (Y-axis)
 Pixel size: max. 512 x 512 pixels
 Scanning speed: 30 fps (512 x 512 pixels) to 230 fps
 (512 x 64 pixels),
 15,600 lines/sec (line speed)
 Zoom: 7 steps (1x, 1.5x, 2x, 3x, 4x, 6x, 8x)
 Scanning mode: X-Y, Line
 Acquisition method: Standard image acquisition,
 High-speed image acquisition, Simultaneous photo activation
 and image acquisition
Dichroic mirror: Low-angle incidence method
 Position: 8
 Standard filter: 405/488, 405/488/561, 405/488/561/638, 405/488/543/638, 457/514, BS20/80
Pinhole: 12-256µm variable (1st image plane)
Spectral detector
Number of channels: 32 channels
Spectral image acquisition speed: 4 fps (256 x 256 pixels), 1000 lps
Maximum wavelength and resolution: 80nm (2.5nm), 192nm (6nm), 320nm (10nm)
 Wavelength range variable in 0.25nm steps
Unmixing: High-speed unmixing, Precision unmixing
Z step: 0.025µm
Compatible microscopes: ECLIPSE Ti-E inverted microscope,
 ECLIPSE TE2000-E inverted microscope,
 ECLIPSE 90i upright microscope,
 ECLIPSE FN1 fixed stage microscope
Option: Motorized XY stage, High-speed Z stage, VAAS
Software
Display/image generation: 2D analysis, 3D volume rendering/orthogonal, 4D analysis, spectral unmixing
Image format: JP2, JPG, TIFF, BMP, GIF, PNG, ND2, JFF, JTF, AVI, ICS/IDS
Application: FRAP, FLIP, FRET, photo activation, three-dimensional time-lapse imaging, multipoint time-lapse imaging, colocalization
Control computer
OS: Microsoft Windows® XP 32bit SP (English version)
CPU: Intel Xeon 5160 (3GHz/1333MHz/dual core) or higher
Memory: 4GB or more
Hard disk: SAS (15,000rpm), 160GB or more x2, RAID 0 configuration
Data transfer: Dedicated data transfer I/F
Monitor: 1600 x 1200 or higher resolution, 2 LCD monitor configuration recommended
Installation condition: Temperature 5 - 35°C, humidity 65% (RH) or less (non-condensing)
Power Source
Controller: Input voltage: 100–240VAC ±10% 50–60Hz
 Current rating: 5A @100VAC
 Overcurrent protection: main breaker 15A
LU-LR 4-laser power source rack: Power source for Ar laser and control circuit: 100VAC,
 15A/115VAC, 15A/230VAC, 7.5A, 50/60Hz (breaker 15A)
 Power source for lasers except Ar laser: 100VAC, 3A/115VAC,
 3A/230VAC, 1.5A, 50/60Hz (breaker 5A)

A1

Input/output port: Laser input port: 3 (FC x2, direct x1)
 Signal output port: 4 (SMA x2, FC x1, VAAS x1)
Laser
Wavelength and power: 405LD: max. 38mW, Multi-Ar (457/488/514): max. 65mW, 488DPSS: max. 75mW, 561DPSS: max. 25mW, 543HeNe: max.
 1mW, 638LD: max. 10mW
 440LD: max. 15mW (available as option)
Modulation: Method: AO (Acousto) device or drive current control
 Control: power control for each wavelength, Return mask, ROI exposure control
Laser unit: Standard: LU4 4-laser unit
 Optional: C-LU3EX 3-laser unit EX
Standard fluorescence detector
Detector: 4 PMT
Filter cube: 6 filter cubes commonly used for a microscope mountable on each of three filter wheels
 Recommended wavelengths: 450/50, 482/35, 525/25, 595/50, 700/75, 540/30, 515/30, 585/65
Diascopic detector
Detector: PMT
Scanning head
Scanning: Standard image acquisition
 Scanner: non-resonant scanner x2
 Pixel size: max. 4096 x 4096 pixels
 Scanning speed: 4 fps (512 x 512 pixels)
 Zoom: 1-1000x continuously variable
 Scanning mode: X-Y, XY rotation, Free line, Line Z
Dichroic mirror: Low-angle incidence method
 Position: 8
 Standard filter: 405/488, 405/488/561, 405/488/561/638, 405/488/543/638, 457/514, BS20/80
Pinhole: 12-256µm variable (1st image plane)
Spectral detector
Number of channels: 32 channels
Spectral image acquisition speed: 4 fps (256 x 256 pixels), 1000 lps
Maximum wavelength and resolution: 80nm (2.5nm), 192nm (6nm), 320nm (10nm)
 Wavelength range variable in 0.25nm steps
Unmixing: High-speed unmixing, Precision unmixing
Z step: 0.025µm
Compatible microscopes: ECLIPSE Ti-E inverted microscope,
 ECLIPSE TE2000-E inverted microscope,
 ECLIPSE 90i upright microscope,
 ECLIPSE FN1 fixed stage microscope
Option: Motorized XY stage, High-speed Z stage, VAAS
Software
Display/image generation: 2D analysis, 3D volume rendering/orthogonal, 4D analysis, spectral unmixing
Image format: JP2, JPG, TIFF, BMP, GIF, PNG, ND2, JFF, JTF, AVI, ICS/IDS
Application: FRAP, FLIP, FRET, photo activation, three-dimensional time-lapse imaging, multipoint time-lapse imaging, colocalization
Control computer
OS: Microsoft Windows® XP 32bit SP (English version)
CPU: Intel Xeon 5160 (3GHz/1333MHz/dual core) or higher
Memory: 4GB or more
Hard disk: SAS (15,000rpm), 160GB or more x2, RAID 0 configuration
Data transfer: Dedicated data transfer I/F
Monitor: 1600 x 1200 or higher resolution, 2 LCD monitor configuration recommended
Installation condition: Temperature 5 - 35°C, humidity 65% (RH) or less (non-condensing)
Power Source
Controller: Input voltage: 100–240VAC ±10% 50–60Hz
 Current rating: 5A @100VAC
 Overcurrent protection: main breaker 15A
LU-LR 4-laser power source rack: Power source for Ar laser and control circuit: 100VAC,
 15A/115VAC, 15A/230VAC, 7.5A, 50/60Hz (breaker 15A)
 Power source for lasers except Ar laser: 100VAC, 3A/115VAC,
 3A/230VAC, 1.5A, 50/60Hz (breaker 5A)