Nikon Instruments / Products / Microscope Systems / Inverted Microscopes / Eclipse TS100/TS100F

Nikon Eclipse TS100 Inverted Microscope

Eclipse TS100/TS100F Specifications

Optical System:  CFI60 infinity system, parfocal distance 60mm
Main body:  Low profile, sturdy main body stand with stage supported on both ends
Focusing:  Vertical objective movement; Coarse stroke: 37.7mm per rotation, Fine stroke: 0.2mm per rotation
Eyepiece Tube:  For TS100: Sidentopf-type binocular tube; For TS100F: Sidentopf type trinocular tube (light distribution, bino/photo: 100/0, 0/100) Interpupillary distance: 50-75mm, Eyepoint height: 400mm from table, Inclination: 45° from horizon
Nosepiece:  Quintuple nosepiece, backward-facing type
Plain stage:  Stage size: 170 x 225mm, Stage height: 195mm from table, Acrylic window provided, Auxiliary stage attachable
Illumination: High luminescent white LED illuminator (Eco-illumination)
6V30W halogen lamp, Built-in heat absorbing filter and diffuser, 2 filters (45-mm NCB11, ND8, GIF) mountable with filter frame
Compliant multi-voltage (100 V-240 V)
Slider:  Non-centerable phase slider (PhL, Ph1, 1 empty position); Centerable phase slider (PhL, Ph1, 1 empty position), Ph2 ring (optional); HMC slider (MCI, MC2, MC3)
Attachable mechanical stage:  Stage movement: 126 x 80mm; Accepts several microplate holders
Holder: Terasaki holder (accepts 65mm petri dish); Slide glass holder (accepts 54mm petri dish); Hemacytometer holder
Filter:  45mm NCB11, ND8 and GIF (green interference)
Eyepiece lens:  C-W 10x (F.O.V. 22mm), C-W 15x (F.O.V. 16mm)
Condenser (without condenser O.D. 190mm):  ELWD condenser: N.A. 0.3 (O.D. 75mm)
Epi-fluorescence attachment:  Field diaphragm, Fluorescence filter block holder (2 filter blocks mountable, 1 empty position), Heat absorbing filter, Lamphouse for 50W mercury lamp, Light shielding plate, UV-cut filter (detachable)

Email Updates

Nikon Instruments Inc. on LinkedIn