Nikon Instruments / Products / Microscope Systems / Confocal Microscopes / A1R MP Multiphoton Confocal

A1R MP Multiphoton Confocal Microscope

A1R MP Specifications

Input/output Port 3 laser input ports
4 signal output ports (for 4-PMT detector, spectral detector, VAAS, optional detector*1)
Laser for Confocal Microscopy  
Compatible Laser

Laser diode (405 nm, max. 36mW), Laser diode (440 nm, max. 20mW), Ar laser (457 nm/478 nm*2/514 nm, max. 65 mW), Solid-state laser (488 nm, max.

20mW), Solid-state laser (561 nm, max. 25mW), G-HeNe laser (543 nm, max. 1 mW), Laser diode (638 nm, max. 10mW)

Modulation Method: AOTF (Acousto-Optical Tunable Filter) or AOM (Acousto-Optical Modulator) device
Control: power control for each wavelength, return mask, ROI exposure control
Laser Unit Standard: 4-laser module or 3-laser module EX
Optional: 3-laser module EX (when 4-laser module is chosen as standard laser unit)
Laser for Multiphoton Microscopy  
Compatible Laser Mai Tai HP DeepSee (Newport Corp.)
Chameleon Vision II (Coherent Inc.)
Modulation Method: AOM (Acousto-Optical Modulator) device
Control: power control, return mask, ROI exposure control
Incident Optics 700-1000 nm, auto alignment
Standard 4-Channel Detector  
Wavelength 400-750 nm (400-650 nm for multiphoton observation)
Detector 4 PMT
Filter Cube 6 filter cubes commonly used for a microscope mountable on each of three filter wheels
Recommended wavelengths for multiphoton/confocal observation: 450/50, 482/35, 515/30, 525/50, 540/30, 585/65, 595/50, 700/75
Diascopic Detector  
Wavelength 440-645 nm
Detector PMT
NDD for Multiphoton Microscopy  
Wavelength 400-650 nm
Detector 4 PMT
Filter Cube Filter cubes commonly used for a microscope
Recommended filter sets for multiphoton: 492SP, 525/50, 575/25, 692/53, DM458, DM495, DM511, DM560, DM593
Unmixing Channel unmixing
Image Bit Depth 4096 gray intensity levels (12 bit)
Scanning Head  
Scanning FOV: Square inscribed in a ∅18 mm circle
Standard image acquisition
Scanner: non-resonant scanner x2
Pixel size: max. 4096 x 4096 pixels
Scanning speed: standard 1 fps (512 x 512 pixels), max. 4 fps (512 x 512 pixels)
Zoom: 1-1000x continuously variable
Scanning mode: X-Y, XY rotation, Free line*3, Line Z
High-speed image acquisition
Scanner: resonant scanner (X-axis, resonance frequency 7.8 kHz), non-resonant scanner (Y-axis)
Pixel size: max. 512 x 512 pixels
Scanning speed: 30 fps (512 x 512 pixels) to 420 fps (512 x 32 pixels),
15,600 lines/sec (line speed)
Zoom: 7 steps (1x, 1.5x, 2x, 3x, 4x, 6x, 8x)
Scanning mode: X-Y, Line
Acquisition method: Standard image acquisition, High-speed image acquisition, Simultaneous photo activation and image acquisition*4
Dichroic Mirror Low-angle incidence method
Position: 8
Standard filter: 405/488, 405/488/561, 405/488/561/638, 457/514, 405/488/543/638, BS20/80, IR, 405/488/561/IR
Pinhole 12-256 μm variable (1st image plane)
Wavelength Detection Range 400 nm-750 nm (400 nm-650 nm with multiphoton microscopy)
Number of Channels 32 channels
Spectral Image Acquisition Speed 4 fps (256 x 256 pixels), 1000 lps
Wavelength Resolution 2.5 nm, 6 nm, 10 nm
Wavelength range variable in 0.25 nm steps
Unmixing High-speed unmixing, Precision unmixing
Compatible Microscopes ECLIPSE Ti-E inverted microscope, ECLIPSE FN1 fixed stage microscope
Z step
Z Step Ti-E: 0.025 μm, FN1 stepping motor: 0.05 μm
Option Motorized XY stage (for Ti-E), High-speed Z stage (for Ti-E), High-speed piezo objective-positioning system (for FN1), VAAS
Software  
Display/Image Generation 2D analysis, 3D volume rendering/orthogonal, 4D analysis, spectral unmixing
Image Format JP2, JPG, TIFF, BMP, GIF, PNG, ND2, JFF, JTF, AVI, ICS/IDS
Application FRAP, FLIP, FRET, photo activation*4, three-dimensional time-lapse imaging, multipoint time-lapse imaging, colocalization
OS Microsoft Windows® Vista Business 64-bit SP1 (Japanese version /English version)
CPU Intel Xeon X5570 (2.98 GHz/8 MB/1333 MHz)
Memory 12 GB
Hard Disk SAS (15,000 rpm), 300 GB x2, RAID 0 configuration
Data Transfer Dedicated data transfer I/F
Monitor 1600 x 1200 or higher resolution, dual monitor configuration recommended
Vibration Isolated Table 1800 (W) x 1500 (D) mm recommended, or 1500 (W) x 1500 (D) mm

AMS en change location

Email Updates